The Oil Test Centre is usually supplied in a robust, NATO approved, metal case (see below), but can also be provided in a portable roller case for onsite use (above).
The Oil Test Centre is supplied with an electronic test console and cells for different test parameters, providing you with a sophisticated range of oil analysis testing equipment.
The Oil Test Centre includes an electronic Viscosity Tester. Configurations are available that include a Heated or Unheated Viscometer . Below: Kittiwake Heated Viscometer
Above: The Oil Test Centre console features an easy-to-use keypad and digital display for simple operation and highly accurate oil test results.
Reagents for all oil tests are supplied. Replacement reagent packs are also available at short notice.
Below: Different configurations of the Oil Test Centre are available (see chart). We also supply the Oil Test Centre in a Portable Roller Case for industrial use.
.jpg)
Home >> Fuel & Oil Testing >> Applications >> Hydraulics & Turbines
Oil Test Centre (FG-K1-100-KW)
The Oil Test Centre is a suite of electronic oil analysis testing equipment designed for use in a very harsh industrial environment. It has been independently correlated against laboratory test procedures and is approved by the Joint Oil Analysis Program.
The Oil Test Centre consists of an electronic console with oil test cells for Insolubles, Total Base Number (TBN), Total Acid Number (TAN) and Water in Oil. Configurations are available with either a Heated Viscometer or Unheated Viscometer. Test equipment is supplied in either a rugged metal case, or a fully portable, samsonite case.
Oil Analysis: Tests
The tests are easy to perform and provide laboratory grade results. The following test equipment is supplied:
Console to PC Data link
A simple piece of software is available with the Oil Test Centre console that allows the data to be downloaded from the console to a PC in comma delimited text format which can be read by standard spreadsheet and data base applications, such as Microsoft Excel and Access.
The downloaded data can be manipulated in any format required, such as lists, graphs or charts for sophisticated analysis of oil trends.
Recommended Configurations
The Oil Test Centre is available with various options covering tests for diesel and gas engines, turbines, hydraulics and gear oils. The equipment supplied can be tailored to your requirements, with configurations are available including the Heated Viscometer. Two ranges of the electronic Water in Oil test are available, as are two different Total Acid Number (TAN) ranges (see chart below for details).
The Oil Test Centre is supplied as standard in a NATO approved, robust metal case. However, portable cases are also available (see photograph, right) - simply enquire for details.
See the chart below for a summary of the main configurations available.
FG-K1-100-KW-A and FG-K1-100-KW-A-H are suitable for diesel engine applications.
FG-K1-100-KW-B and FG-K1-100-KW-B-H are suitable for steam turbines, compressors & Gearboxes.
FG-K1-100-KW-C, FG-K1-100-KW-C-H, FG-K1-100-MO-S
and FG-K1-100-MO-S-H are suitable for aviation lubes & hydraulics.
MILITARY STOCK CODE
Oil Test Centre, NATO Stock Number NSN:6630-99-215-5830
Click here to view the Oil Test Centre Instruction Manual
Dimensions: | 57.5 x 36 x 31 cms |
Operating Temperature: | 10 - 50°C |
Power: | 110/240 VAC (autoselected) |
Weight: | 18kg |
Want to know more?
Contact Kittiwake here
Download PDF of Oil Test Centre 
| Recommended Configurations | |||||||||||||||||||
Part Number |
Tests Included: |
||||||||||||||||||
Water |
TBN |
TAN |
Viscosity |
Insolubles |
|||||||||||||||
FG-K1-100-KW-A |
0-2.5% | 5 - 99 | Unheated | 0-3.5% | |||||||||||||||
FG-K1-100-KW-A-H |
0-2.5% | 5 - 99 | Heated | 0-3.5% | |||||||||||||||
FG-K1-100-KW-B |
0-2.5% | 0-6 | Unheated | ||||||||||||||||
FG-K1-100-KW-B-H |
0-2.5% | 0-6 | Heated | ||||||||||||||||
FG-K1-100-KW-C |
0-6000 ppm | 0-6 | Unheated | 0-3.5% | |||||||||||||||
FG-K1-100-KW-C-H |
0-6000 ppm | 0-6 | Heated | 0-3.5% | |||||||||||||||
FG-K1-100-MO-S |
0-6000 ppm | 0-6 | Unheated | 0-3.5% | |||||||||||||||
FG-K1-100-MO-S-H |
0-6000 ppm | 0-6 | Heated | 0-3.5% | |||||||||||||||
FG-K1-100-KW-D |
0-6000 ppm | 0-3 | Unheated | ||||||||||||||||
FG-K1-100-KW-D-H |
0-6000 ppm | 0-3 | Heated | ||||||||||||||||
FG-K1-100-KW-E |
0-6000 ppm | 5 - 99 | 0-3 | Unheated | 0-3.5% | ||||||||||||||
FG-K1-100-KW-E-H |
0-6000 ppm | 5 - 99 | 0-3 | Heated | 0-3.5% | ||||||||||||||
FG-K1-100-KW-F |
0-6000 ppm | 5 - 99 | Unheated | 0-3.5% | |||||||||||||||
FG-K1-100-KW-F-H |
0-6000 ppm | 5 - 99 | Heated | 0-3.5% | |||||||||||||||
| Overview of Test Specifications | |||||||||||||||||||
Test |
Range |
Calculations |
Typical Accuracy |
Correlation |
|||||||||||||||
Viscosity (heated) |
15-810 cSt at 40°C and 50°C. RMA 10 to RML 55. SAE 5 to 50 | cSt at 100°C. CCAI. Density 50 to 15°C in vacuo | +/- 3% (15-450 cSt) | ASTM D445/ IP71 | |||||||||||||||
Viscosity (unheated) |
15-500 cSt at 40°C.SAE 5 to 50 | cSt at 100°C | +/- 2% (15-320 cSt) | ASTM D445/ 1P71 | |||||||||||||||
Water |
0 - 2.5% w/w. or 0-6000 ppm | +/- 0.1% or +/- 100 ppm | ASTM D4928/ IP386 | ||||||||||||||||
TBN |
5-99 mg/KOH | +/- 5% | IP 400 | ||||||||||||||||
Insolubles |
0 - 3.5% w/w | +/- 0.1% | IP 316 Mobil Soot Index | ||||||||||||||||
TAN |
0-6 mg KOH | +/- 0.2 TAN | ASTM D 664/ IP 177, SAE ARP 5088 (Modified IP139, ASTM D 974) | ||||||||||||||||
| Spares for Test Equipment | |||||||||||||||||||
Equipment |
Part Number |
||||||||||||||||||
Water in Oil Cell 0-2.5% |
AS-K1-104 | ||||||||||||||||||
Water in Oil Cell 0-6000ppm |
AS-K1-105 | ||||||||||||||||||
TBN Cell 0-40 TBN |
AS-K1-106 | ||||||||||||||||||
Insolubles Cell 0-3.5% |
AS-K1-107 | ||||||||||||||||||
TAN Cell 0-6 TAN |
AS-K1-108 | ||||||||||||||||||
Viscometer (heated) |
FG-K1-200-KW | ||||||||||||||||||
Viscometer (unheated) |
AS-K1-110 | ||||||||||||||||||
Spare Seals, Balls and Fuses |
FG-K10503-KW | ||||||||||||||||||
Viscometer End Cap |
AS-11097 | ||||||||||||||||||
Viscometer End Plug |
AS-11098 | ||||||||||||||||||
OTC Data Download Cable |
AS-K12219 | ||||||||||||||||||
OTC Data Download Software |
MA-13136-KW | ||||||||||||||||||
Fuel and Lube oil Training Manual |
MA-K10598 | ||||||||||||||||||
Water in Oil Reagent Pack |
FG-K2-001-KW | ||||||||||||||||||
Total Base Number Reagent Pack |
FG-K2-002-KW | ||||||||||||||||||
Insolubles Reagent Pack |
FG-K2-003-KW | ||||||||||||||||||
Water/Insolubles/Total Base Number Reagent Pack |
FG-K2-004-KW | ||||||||||||||||||
Insolubles in Lubricating Oil Reagent Pack |
FG-K1-006-KW | ||||||||||||||||||
Salt Water Contamination Kit |
FG-K1-005-KW | ||||||||||||||||||
Total Acid Number (0 - 3) Reagent Pack |
FG-K2-006-KW | ||||||||||||||||||
Total Acid Number (0 - 6) Reagent Pack |
FG-K2-005-KW | ||||||||||||||||||