Oil Test Centre (FG-K1-100-KW)

The Oil Test Centre is a suite of electronic oil analysis testing equipment designed for use in a very harsh industrial environment. It has been independently correlated against laboratory test procedures and is approved by the Joint Oil Analysis Program.

The Oil Test Centre consists of an electronic console with oil test cells for Insolubles, Total Base Number (TBN), Total Acid Number (TAN) and Water in Oil. Configurations are available with either a Heated Viscometer or Unheated Viscometer. Test equipment is supplied in either a rugged metal case, or a fully portable, samsonite case.

Oil Analysis: Tests

The tests are easy to perform and provide laboratory grade results. The following test equipment is supplied:

Viscosity
(electronic) for fuel & lube oils




Insolubles
(electronic)


Water in Oil
(electronic)

Features

Joint Oil Analysis Program Approval. Used for On-Ship Lube Oil Analysis Programs in UK Navy, US Navy (LANT fleet PAC fleet, MSC), Spanish Navy, Australian Navy, French Navy, Brunei Navy, Italian Navy


Data can be easily downloaded to your PC for accurate and sophisticated analysis of trends and readings.


A self contained suite of equipment, purpose designed for reliable and accurate oil condition monitoring.


Housed in its own robust metal case - designed for use in harsh environments. Very easy to clean and store.


Easy and quick to use - suitable for non-technical personnel. Comes complete with a comprehensive instruction manual and training CD.


Complements the Kittiwake range of sampling equipment. Replacement reagents or additional test equipment easily available and can be shipped at short notice.



Console to PC Data link

A simple piece of software is available with the Oil Test Centre console that allows the data to be downloaded from the console to a PC in comma delimited text format which can be read by standard spreadsheet and data base applications, such as Microsoft Excel and Access.

The downloaded data can be manipulated in any format required, such as lists, graphs or charts for sophisticated analysis of oil trends.


Recommended Configurations

The Oil Test Centre is available with various options covering tests for diesel and gas engines, turbines, hydraulics and gear oils. The equipment supplied can be tailored to your requirements, with configurations are available including the Heated Viscometer. Two ranges of the electronic Water in Oil test are available, as are two different Total Acid Number (TAN) ranges (see chart below for details).

The Oil Test Centre is supplied as standard in a NATO approved, robust metal case. However, portable cases are also available (see photograph, right) - simply enquire for details.

See the chart below for a summary of the main configurations available.

FG-K1-100-KW-A and FG-K1-100-KW-A-H are suitable for diesel engine applications.

FG-K1-100-KW-B and FG-K1-100-KW-B-H are suitable for steam turbines, compressors & Gearboxes.

FG-K1-100-KW-C, FG-K1-100-KW-C-H, FG-K1-100-MO-S
and FG-K1-100-MO-S-H are suitable for aviation lubes & hydraulics.


MILITARY STOCK CODE
Oil Test Centre, NATO Stock Number NSN:6630-99-215-5830

Click here to view the Oil Test Centre Instruction Manual


Specifications


Dimensions:

57.5 x 36 x 31 cms

Operating Temperature:

10 - 50°C

Power:

110/240 VAC (autoselected)

Weight:

18kg


Want to know more?
Contact Kittiwake here


Download PDF of Oil Test Centre  Yellow Arrow


To view the document you will need Adobe Acrobat Reader installed on your system. If you do not have it click here to proceed to Adobe's website to download it.



Recommended Configurations
Part Number
Tests Included:
Water
TBN
TAN
Viscosity
Insolubles
FG-K1-100-KW-A
0-2.5% 5 - 99 Unheated 0-3.5%
FG-K1-100-KW-A-H
0-2.5% 5 - 99 Heated 0-3.5%
FG-K1-100-KW-B
0-2.5% 0-6 Unheated
FG-K1-100-KW-B-H
0-2.5% 0-6 Heated
FG-K1-100-KW-C
0-6000 ppm 0-6 Unheated 0-3.5%
FG-K1-100-KW-C-H
0-6000 ppm 0-6 Heated 0-3.5%
FG-K1-100-MO-S
0-6000 ppm 0-6 Unheated 0-3.5%
FG-K1-100-MO-S-H
0-6000 ppm 0-6 Heated 0-3.5%
FG-K1-100-KW-D
0-6000 ppm 0-3 Unheated
FG-K1-100-KW-D-H
0-6000 ppm 0-3 Heated
FG-K1-100-KW-E
0-6000 ppm 5 - 99 0-3 Unheated 0-3.5%
FG-K1-100-KW-E-H
0-6000 ppm 5 - 99 0-3 Heated 0-3.5%
FG-K1-100-KW-F
0-6000 ppm 5 - 99 Unheated 0-3.5%
FG-K1-100-KW-F-H
0-6000 ppm 5 - 99 Heated 0-3.5%

Overview of Test Specifications
Test
Range
Calculations
Typical Accuracy
Correlation
Viscosity (heated)
15-810 cSt at 40°C and 50°C. RMA 10 to RML 55. SAE 5 to 50 cSt at 100°C. CCAI. Density 50 to 15°C in vacuo +/- 3% (15-450 cSt) ASTM D445/ IP71
Viscosity (unheated)
15-500 cSt at 40°C.SAE 5 to 50 cSt at 100°C +/- 2% (15-320 cSt) ASTM D445/ 1P71
Water
0 - 2.5% w/w. or 0-6000 ppm +/- 0.1% or +/- 100 ppm ASTM D4928/ IP386
TBN
5-99 mg/KOH +/- 5% IP 400
Insolubles
0 - 3.5% w/w +/- 0.1% IP 316 Mobil Soot Index
TAN
0-6 mg KOH +/- 0.2 TAN ASTM D 664/ IP 177, SAE ARP 5088 (Modified IP139, ASTM D 974)

Spares for Test Equipment
Equipment
Part Number
Water in Oil Cell 0-2.5%
AS-K1-104
Water in Oil Cell 0-6000ppm
AS-K1-105
TBN Cell 0-40 TBN
AS-K1-106
Insolubles Cell 0-3.5%
AS-K1-107
TAN Cell 0-6 TAN
AS-K1-108
Viscometer (heated)
FG-K1-200-KW
Viscometer (unheated)
AS-K1-110
Spare Seals, Balls and Fuses
FG-K10503-KW
Viscometer End Cap
AS-11097
Viscometer End Plug
AS-11098
OTC Data Download Cable
AS-K12219
OTC Data Download Software
MA-13136-KW
Fuel and Lube oil Training Manual
MA-K10598
Water in Oil Reagent Pack
FG-K2-001-KW
Total Base Number Reagent Pack
FG-K2-002-KW
Insolubles Reagent Pack
FG-K2-003-KW
Water/Insolubles/Total Base Number Reagent Pack
FG-K2-004-KW
Insolubles in Lubricating Oil Reagent Pack
FG-K1-006-KW
Salt Water Contamination Kit
FG-K1-005-KW
Total Acid Number (0 - 3) Reagent Pack
FG-K2-006-KW
Total Acid Number (0 - 6) Reagent Pack
FG-K2-005-KW