Kittiwake's Cloud Point Detector is supplied to the military in a robust metal case, but can also be provided in the standard Kittiwake portable carry case.

Home >> Fuel & Oil Testing >> Equipment >> Fuel & Lube Oil Laboratories
FG-K12663-KW: Cloud Point Detector (Standard Range)
For testing cloud point in the range from +20 °C to -5°C (Diesel conforming to NATO F-76 std).
Standard Cloud Point Detector housed in an industrial, metal case (see photograph opposite).
MILITARY STOCK CODE
Oil Test Centre, NATO Stock Number NSN:6630-99-215-5830
FG-K16954-KW: Cloud Point Detector (Extended Range)
For testing extended range cloud points from +25 °C to -20 °C
Extended Range Cloud Point Detector housed in a portable carry case.
Cloud Point Measurement
Kittiwake's Cloud Point Detectors are automatic devices, used for detecting the temperature at which wax crystals form in a translucent oil sample. This causess the oil to become cloudy in appearance, hence the term "cloud point".
The measurement of cloud point is of importance to the marine industry, particularly in climates with low ambient temperatures. If the temperature drops below the cloud point, wax crystals forming in the fuel may block filters and cause fuel delivery problems.
This device operates by cooling an oil sample and monitoring the intensity of light transmitted through it. The unit allows cloud point tests to be carried out before accepting the fuel and a normal test takes less than 10 minutes.
Cloud Point Detector Features
Ambient Temperature: | +20°C to +30°C |
Diesel Test Temperature: | Room Temperature |
Dimensions (Boxed): | 575 x 360 x 310mm |
Dimensions (Unit): | 250 x 255 x 115mm |
Display: | 124x64 graphics LCD with LED backlight |
Environmental protection: | IP21 |
Example test time: | (Using Range 2) -1°C cloud point detected in <6 minutes |
Frequency: | 50/60Hz |
Power: | 100-250 VAC |
Range - FG-K12663-KW: | -5°C to +20°C Diesel conforming to NATO F-76 std |
Ranges - FG-K16954-KW: | User selectable: +25°C to 0°C, +15°C to -10°C, +5°C to -20°C |
Sample size: | 0.5ml |
Weight: | Unit alone 2.6kg, Kit in metal case 11kg, Kit in portable case 6kg |